LS195 Glossometer Intelligent Stone Metal Luster Industrial Gloss Testing Device
Specifications:
Item Type: Glossy Meter
Model: LS195
Material: ABS.
Battery: 1 x Rechargeable Lithium Battery 3.7V 650mAh (Battery Included)
Purpose: Sampling and gloss measurement comparison
Application: paint, coating, ink, plastic, paper, ceramic tile, ceramic, stone, metal, electroplating layers
Measuring angle: 60°
Measuring spot: Approx. 9 x 15mm / 0.4 x 0.6in
Minimum size of test object: approx. 0.8 x 0.4 inches (20 x 10 mm)
Measurement Range: 0-200GU
Resolution: 0.1GU
Reproductivity:
0-100GU: ±0.2GU
100-200GU: ±0.2% of reading value
Reproductivity:
0-100GU: ±0.5GU
100-200GU: ±0.5% of reading value
Error: 0.1GU
Display error:
0-100GU: ±1.5GU
100-200GU: ±1.5% of reading value
System language: Chinese, English
Display: 9.4 x 128 dot LCD
Charging Interface: USB (Type C)
Data Interface: USB
Operating Temperature Range: 10 ~ 45, 0 ~ 85% RH (without condensation)
Storage temperature range: -10 - 60, 0 - 85% RH (non-condensation).
How to use:
1. Power on/off
Power On: Short press the device button to turn on when it is off.
Shutdown: Press and hold the device button to shut down. If there is no operation for more than 3 minutes, the device will automatically shut down.
Mechanical
2. Parameter setting
In shutdown state, press and hold the device button for 3 seconds to enter parameter setting mode.
A. Language (language)
Short press the button to select Chinese or English, and long press the button on the 3S to complete the setting, then enter the next setting.
B. Select Mode
Short press the button to select normal mode or QC mode, and long press the button of the 3S to complete the setting, finish setting and enter the calibration interface.
3. Device calibration
Place the device on the standard board and turn on the opportunity to enter the calibration interface. The user can perform calibration operations according to the calibration prompt, and it will enter the measurement interface once the calibration is completed. Turn on the device without placing it on the standard board, the calibration will skip and input directly into the measurement interface. If you see a message that indicates the calibration failure, then there is a reason you can think of:
Standard boards are dirty Clean the standard board with a special lens cloth and buckle the host to the base.
The host is not near the standard board. Simply buckle it to the base again.
If the ambient temperature changes very well, please make sure the temperature and ambient temperature are close and there is no condensation on the lens or base reference board in the device test port before resuming measurement.
The attenuation of the light source will not work properly, and you need to return it to the factory for inspection and repairing.
If the device has a "failed calibration" message, long press the button to skip the calibration.
Package List:
1 x Glossometer (battery included).
1 x U Disk
D x 1