Introduction:

A: A sample used for measuring very low viscosity

B: Sample size is 25ml enough

C:Cylindrical rotor can accurately calculate shear speed so the detailed analysis of the product is easy.


Technical Specifications

Measurement range: 1-100 (MPa S)

Accuracy: 0.02


0 Rotor Ultra Low Viscosity Adapter, How to use:

1, No. Less than 15 mpa.s of the liquid selection of 0 rotors, No. 0 rotors without protective frame, No. 0 rotors attached to connection screw (counterclockwise rotation installed)

2, Fix the fixed sleeve to the bottom of the device cylinder, and the sleeve fixing screw.

3 Pour 20-25 ml of measuring liquid into an external test cylinder tube (with bottom)

4 Insert the outer test tube into the fixed sleeve and tighten with the test tube fixing screw. Need to pay for the cone end of the test tube fixing screw screwed into the triangular groove of the top end of the test cylinder (With the circular hole on the sides, the hole will be observed if the triangle of the test tube is in the center of the circular center) Test can start after the liquid temperature is controlled

5, In the bottom free test tube, the method is basically the same (except 3), dip into the groove ring of the test cylinder as an external test tube and rotor fluid line.