The Rotor 0 Ultra Low Viscosity Adapter is the following:
1.For liquid less than 15MPa.s, choose rotor 0, use rotor 0 without protective frame, and attach rotor 0 to the connection screw (rotate counterclockwise).
2. Slide the fixing sleeve into the lower cylinder of the device, tighten with the socket fixing screw.
Pour 3.20-25ml test fluid into the outer test cylinder (bottom).
4. Sleeve the outer test cylinder in the fixed sleeve, tighten with the test cylinder fixing screw, pay attention to the cone end of the test cylinder fixing screw,
Threaded into triangular grooves on the top of the outer test cylinder (you can observe whether the triangular groove of the test cylinder is centered in the round hole on the side) After you control the temperature of the liquid to measure, you can perform a test.
5. If you use a bottom free outer test cylinder, the method of dipping the outer test cylinder and rotor into the groove ring of the external test cylinder as a liquid line (except 3).