Feature

●Suitable for NDJ-1 / 5S / 8S / 9S Digital Viscosity Meter
●Rotor 0 can measure the low viscosity of 0.1 mPa.s. Specially designed to measure samples of shallow viscosity Sample volume is only 0.9 fl oz (25 ml). Cylindrical rotor accurately calculates shear speed for detailed product analysis
●A complete set of precision technology made of 304 stainless steel, stable performance, corrosion resistance, and high temperature resistance
●Micro bottom sleeves with a sample volume of only 25ml
●Comes standard with rotor protection sleeve with liquid level markings Measurement is most often performed when the liquid is parallel to the dot marker. The No.0 rotor can measure the viscosity range of 0.1 to 100 MPa.s, and the viscosity standard for regular tap water is 0.8 to 1.5 MPa.s


Description

The Rotor 0 Ultra Low Viscosity Adapter is the following:
1.For liquid less than 15MPa.s, choose rotor 0, use rotor 0 without protective frame, and attach rotor 0 to the connection screw (rotate counterclockwise).
2. Slide the fixing sleeve into the lower cylinder of the device, tighten with the socket fixing screw.
Pour 3.20-25ml test fluid into the outer test cylinder (bottom).
4. Sleeve the outer test cylinder in the fixed sleeve, tighten with the test cylinder fixing screw, pay attention to the cone end of the test cylinder fixing screw,
Threaded into triangular grooves on the top of the outer test cylinder (you can observe whether the triangular groove of the test cylinder is centered in the round hole on the side) After you control the temperature of the liquid to measure, you can perform a test.
5. If you use a bottom free outer test cylinder, the method of dipping the outer test cylinder and rotor into the groove ring of the external test cylinder as a liquid line (except 3).