Feature

●Compatible with NDJ-8SNDJ-9S rotational viscosity meter NDJ-5S NDJ-9S
●No. 0 Rotor Fluid Selection Less than 15 mpa.s
●Sample used to measure very low viscosity
●Cylindrical rotor can accurately calculate shear speed so it is easy to analyze the product in detail
●No. 0 rotor liquid selection less than 15mpa.s


Description

Product Description

Processor:

A: Samples used to measure very low viscosity

B: Sample size only needs 25ml

C:Cylindrical rotor can accurately calculate shear speed so it is easy to analyze the product in detail.

Technical Specifications

Measurement range: 1-100 (MPa. S)

Accuracy: 0.02

0 Rotor Ultra Low Viscosity Adapter, How to Use

1, Less than 15 mpa.s of the #0 rotor of the liquid selection, No.0 rotor, without protective frame, Rotor No.0 is attached to the connection screw (counterclockwise rotation)

2, Fix the fixed sleeve to the bottom of the instrument cylinder and screw it with the sleeve fixing screw.

3. Pour 20-25ml measuring fluid into the outside test cylinder tube (with bottom).

4. Insert the outer test tube into the fixed sleeve and tighten with the test tube fixing screw.When fastening, you must be aware that the conical end of the test tube fixing screw is screwed into the triangular groove of the top end. Test cylinder (round hole on the side to observe whether the triangle of the test tube is centered on the round hole). After you control the temperature of the liquid and start testing.

5. As shown on bottomless test tube, the method of submerging the external test tube and rotor into the groove ring of the test cylinder as a liquid line (except 3).